| Number of Inputs: |
1 (9 pin Redel) |
| Pressure Output: |
2.3 Volt/100 mmHg |
| Power Supply Out: |
+5VDC, +/- 12 VDC |
| Power Supply In: |
100-250 Vac, 50-60 Hz |
| Dimensions: |
H : 6.9 cm (2.7”)
W : 22.6 cm (8.9")
L : 16.8 cm (6.6") |
| Calibration Controls: |
Pressure: 0 mmHg (-2.86V) and 100 mmHg (-0.57V)
Volume (Span 0V to +5V):
- Mouse = 0 uL - 150 uL,
- Rat = 0 uL - 1000 uL
- Rabbit = 0 uL - 2000 uL
- Large = 0 mL - 200 mL
Phase: 0° to 20° (Span 0V to +5V)
Magnitude (Span 0V to +5V):
- Mouse = 0 uS - 5000 uS
- Rat = 0 uS - 5000 uS
- Rabbit = 0 uS - 15000 uS
- Large = 0 mS - 50 mS
|
| Catheter Cable: |
122 cm (48"/ 4 feet) cable for small animal
366 cm (144"/ 12 feet) cable for large animal |
| Output Connector: |
BNC |
| Signal Gain: |
1143 (Pressure) |
| |
|
- hemodynamics
- myocardial infarction
- ischemia/reperfusion injury
- hypertension
- cardiac contractile function
- molecular cardiology
- remodelling
- valve mechanics
- trans-aortic constriction
- diabetes mellitus
- oxidative stress
- vascular disease
- gene therapy
- diastolic funcion
- cardiomyopathy
- stem cell therapy
|